This page is a sub-page to
LOW NOISE AMPLIFIERS. HOW TO OPTIMIZE AND MEASURE PERFORMANCE.
Tested amplifiers.
Measurement system.Linrad was running with the SDR-IP at a bandwidth of 1.2 MHz. An SBL1 Schottky diode mixer was connected to the input with a HP8657A as the LO at 134 MHz. In front of the mixer a filter with more than 60 dB attenuation on the mirror frequency 124 MHz was inserted to ensure that all the signal reaching the SDR-IP on 10 MHz originated in 144 MHz. Low noise amplifiers in front of the filter provide a system NF of 3.4 dB. Another HP8657A signal generator was connected to precision attenuators to provide an accurate match to 50 ohms as measured on the connector used to connect to the different amplifiers under test.Measurement conditions.All measurements on this page were made on Aug 13 2012. This day with stable weather the air pressure was reported to be 1024 hPa here in Eskilstuna. The table in Handbook of Chemistry and Physics gives the temperature of boiling water for pressures in mm Hg. 1024 hPa converts to 768.0 mmHg. My lab is 15 meters above sea level so the pressure in my lab should be 766.6 mm Hg. The handbook tells me that pure water would boil at 100.606 °C. For the experiment "battery water" was used. It is demineralized water with a conductivity of less than 5 uS/cm. Hopefully this water is pure enough to boil and freeze at the expected temperatures 0.00 and 100.61 °C which converts to 273.15 K and 373.76 K.Screen dumps. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
MGF1425old. The Y-factor is 1.195 dB. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
MGF1801. The Y-factor is 1.272 dB. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
MGF1425. The Y-factor is 1.272 dB. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||
ATF33143. The Y-factor is 1.293 dB. | |||||||||||||||||||||||||||||||||||||||||||||||||||
2xATF33143. The Y-factor is 1.263 dB. | |||||||||||||||||||||||||||||||||||||||||||||
ATF33143negimp. The Y-factor is 1.297 dB. | |||||||||||||||||||||||||||||||||||||||
PSA4-5043. The Y-factor is 1.193 dB. | |||||||||||||||||||||||||||||||||
FHX05FA/LG. The Y-factor is 1.293 dB. | |||||||||||||||||||||||||||
NONE. The Y-factor is 0.668 dB. | |||||||||||||||||||||
Evaluation.
The total noise is the sum of the thermal noise of the attenuator, the
noise from the tested amplifier and the noise from subsequent stages in the
system 2nd Tinp.
The Y-factor is the ratio of two such sums like this:
| |||||||||||||||||||
Unit Y-factor Y-factor Gain 2nd Tinp TLNA NF (name) (dB) (lin.pow) (dB) (K) (K) (dB) ATF33143negimp 1.297 1.34803 32.8 0.17 15.76 0.230 FHX05FA/LG 1.293 1.34679 34.1 0.13 16.84 0.245 ATF33143 1.293 1.34679 27.4 0.60 16.37 0.238 MGF1801 1.272 1.34029 28.0 0.53 21.98 0.317 MGF1425 1.272 1.34029 30.8 0.28 22.23 0.321 2xATF33143 1.263 1.33752 28.5 0.47 24.47 0.352 MGF1425old 1.195 1.31674 31.2 0.25 44.24 0.617 PSA4-5043 1.193 1.31613 23.8 1.38 43.73 0.610 NONE 0.668 1.16627 0.0 332Table 1.. | |||||||||||||||||
The results in table 1 should rank all the amplifiers properly. There are however some systematic errors that should be taken into account. The cable from the attenuator to the test object is 80 mm of Suhner 02250 high temperature teflon cable (as measured from the copper tube responsible for water-proofing.) The nominal loss is 0.33 dB/m at 150 MHz so this cable would add 0.026 dB to the NF result if it were kept at room temperature over all the 80 mms of its length. It seems reasonable to assume it adds 0.015 dB to the result. In this experiment, a second cable, 225 mm of RG223U was also inserted. The nominal attenuation is 0.16 dB/m so this cable contributes 0.036 dB. The correction for cable losses should be about -0.05 dB. Table 2 shows the corrected results from table 1 and a comparison with a previous comparison of these amplifiers with Linrad used as a NF meter. (That measurement only gives a relative dB scale and the zero point was a guess based on a previous hot/cold measurement.) | |||||||||||||||
Unit NF(table 1) NF(previous) Diff Error NF(prev.corrected) (name) (dB) (dB) (dB) (dB) (dB) ATF33143negimp 0.180 0.205 0.025 -0.009 0.171 FHX05FA/LG 0.195 0.240 0.045 0.011 0.206 ATF33143 0.188 0.271 0.083 0.049 0.237 MGF1801 0.267 0.304 0.037 0.003 0.270 MGF1425 0.271 0.266 -0.005 0.039 0.232 2xATF33143 0.302 0.327 0.025 -0.009 0.293 MGF1425old 0.567 0.557 -0.010 -0.044 0.523 PSA4-5043 0.560 0.631 0.071 0.037 0.597 avg= 0.034Table 2. The relative NF values (previous) in table 2 have to be shifted by the average difference 0.034 dB downwards. That seems to have been the error in the zero point of the relative NF measurements. . | |||||||||||||
Table 2 compares measurements obtained by two different methods. The biggest differences are just below 0.05 dB which means that the error in individual measurements is below 0.03 dB but there is an unknown systematic error that is equal for all amplifiers that is related to errors in the noise temperatures due to cable losses. That error is probably very small. The attenuator is made with standard surface mount resistors. The attenuation varies with temperature. Figure 8 shows the change with temperature. | |||||||||||
Fig. 8. Attenuation vs temperature. The instrument is calibrated to show zero transmission loss with the attenuator at 0 °C. The transmission loss changes by something in the order of 0.01 dB when the attenuator is heated to 100 °C. | |||||||
It is not obvious how to correct the NF results by use of the data from figure 8. Not only the attenuation changes with the temperature. Also the impedance changes a little. There may be a small systematic error here. If the increased loss is due to changed matching to the network analyzer there should be no correction to the NF results, but if the 0.01 dB signal loss is resistive the NF values in table 2 would be too good by 0.01 dB. There could be both phenomena with opposite signs so there is an unknown systematic error here, but probably it is not larger than 0.01 dB. The amplifiers on this page were brought to the 15th INTERNATIONAL EME CONFERENCE and measured there in several ways. This link results from EME 2012. shows the raw data as well as the conclusions I think one can draw from a comparison with the data in table 2. (Table 2 was published before the meeting.) Impedance variations.Figure 9 shows the impedance change between 0 and 100 °C. The change is 0.7 ohms. | |||||
Fig. 9. Impedance vs temperature. The instrument is calibrated to show 50 ohms with the attenuator at 0 °C. The impedance change at 144 MHz is 0.7 ohms. | |